| PROBLEM |
CAUSE |
SOLUTION |
Control unit
"ON"
No display |
Interrupted Power
Faulty Data Card |
- Check wall outlet for power
- Check installation of power cord to unit
- Check fuses on back of panel and replace if necessary
- Replace the data card
|
Display reads:
".." |
Indicates completion
of COARSE or FINE mode stimulation |
- Press either button on probe handle to return to standby
mode
|
Display reads:
" EP" |
No Probe Tip
Detected |
- Remove and re-insert probe tip
- Use only new probe tip
- Remove and reinsert probe handle cable into front of unit
|
Display reads:
" EH" |
Tumescence Signal
Too High |
- Penile circumference is greater than 15 cm
- Check penile sensor for twists or kinks
- Check white cable connections
- If problem persists, apply new tumescence sensor and repeat
|
Display reads:
" EL" |
Tumescence Signal
Too High |
- Check white cable connections
- If problem persists, apply new tumescence sensor and repeat
|
Display reads:
" E2" |
Clock Self Check
Error |
- Turn Unit "OFF" and "ON"
- If persists: Do not use unit and call Blue Torch
|
Display reads:
" E3" |
Probe Tip Error |
- Check probe cable connections
- Re-insert probe tip into handle
- Turn Unit "OFF" and "ON"
- If persists, replace the probe tip
|
Display reads:
" E4" |
Data Card Storage
Error
Data Card is full |
- Power off
- Install new data card
- Turn unit "ON"
- If the intent is NOT to use a data cars, cancel the alarm by
simultaneously pressing the Manual Adjustment buttons
|
Display reads:
" E5" |
Power Supply |
- Turn Unit "OFF" and "ON"
- If persists: Do not use unit and call Blue Torch
|
Display reads:
" E7" |
CRC Self Check Error |
- Turn Unit "OFF" and "ON"
- If persists: Do not use unit and call Blue Torch
|
Display reads:
" E8" |
Probe Tip Use
Greater Than 8 Hours |
- Do not reuse probe tip
- Replace with new probe tip
|
Display reads:
" PL" |
Stimulation Circuit
Interrupted |
- Ensure all electrode are touching tissue
- Ensure black lead is clipped to retractor
- Check black lead connection to control unit
- Clear alarm by pressing any button on probe tip
- If persists, replace probe tip
|
| Control Unit Not
Responding to Probe |
Probe Handle
Function |
- Try both buttons
- Check cable connections
- Turn Unit "OFF" then "ON", lights should
flash on probe handle briefly
- If persists, replace probe handle
|
| Intermitent
Stimulation |
Probe Handle
Function |
- Do not hold down probe handle buttons
- Check probe handle connections
|
| No Tumescence
Response When Expected |
System Not
Functioning Properly
No Nerve
Penile Circumference is less than 8 cm
Patient's Current Threshold Low |
- Test unit on abdominal muscle to verify function
- Using COARSE mode allow system to automatically increase
current levels to 20 mA
- Try another location for the Cavermosal nerve
|