Troubleshooting & Setup Guide

CaverMap™ Surgical Aid Set-up Guide


1.)  Pre-operative

  • Open sterile components on back table: Probe Handle Black Patient Return Line White Tumescence Sensor and Cable and White Probe Tip
  • Place CaverMap™ control unit at side and lower half of operating table and visible to surgeon
  • Plug in unit, do not power on until ready to use.
  • Apply Tumescence Gauge to the base of the penis after skin prep, draping and immediately prior to insertion of the Foley catheter
Caution. Penile Gauge contains 0.16 grams of mercury. Do not over-stretch during application In the event of a spill, remove mercury to a covered container and wash exposed skin with soap and water. Follow your institution’s biohazard waste protocol for mercury.

2.) Immediately Prior to Prostate Dissection and From the Sterile Field.

  • Attach the white tumescence cable to the penile gauge and clip to a drape 10 prevent tension.
  • Clip the black Patient return lead to a wound retractor and hand off the connector end for insertion in the control unit.
  • Attach the probe tip to the probe handle and pass off the capped 2nd for insertion into the control unit..

3.) Turn on the unit

4.) It is recommended that the surgeon test the probe tip on observed muscle tissue prior to use.


CaverMap™ Troubleshooting Guide


If the solutions do not resolve the problem, turn the unit "OFF" and then "ON".  If the problem persists:

  • Do not use the CaverMap™ Surgical Aid.

  • Record the error codes.

  • Record the serial number located on the rear panel of the unit and on the probe handle and the lot number of the disposable cables.

  • Call Blue Torch Customer Support at (508)231-1080.

PROBLEM CAUSE SOLUTION
Control unit "ON"
   No display
Interrupted Power
Faulty Data Card
  • Check wall outlet for power
  • Check installation of power cord to unit
  • Check fuses on back of panel and replace if necessary
  • Replace the data card
Display reads:
   ".."
Indicates completion of COARSE or FINE mode stimulation
  • Press either button on probe handle to return to standby mode
Display reads:
   " EP"
No Probe Tip Detected
  • Remove and re-insert probe tip
  • Use only new probe tip
  • Remove and reinsert probe handle cable into front of unit
Display reads:
   " EH"
Tumescence Signal Too High
  • Penile circumference is greater than 15 cm
  • Check penile sensor for twists or kinks
  • Check white cable connections
  • If problem persists, apply new tumescence sensor and repeat
Display reads:
   " EL"
Tumescence Signal Too High
  • Check white cable connections
  • If problem persists, apply new tumescence sensor and repeat
Display reads:
   " E2"
Clock Self Check Error
  • Turn Unit "OFF" and "ON"
  • If persists: Do not use unit and call Blue Torch
Display reads:
   " E3"
Probe Tip Error
  • Check probe cable connections
  • Re-insert probe tip into handle
  • Turn Unit "OFF" and "ON"
  • If persists, replace the probe tip
Display reads:
   " E4"
Data Card Storage Error
Data Card is full
  • Power off
  • Install new data card
  • Turn unit "ON"
  • If the intent is NOT to use a data cars, cancel the alarm by simultaneously pressing the Manual Adjustment buttons
Display reads:
   " E5"
Power Supply
  • Turn Unit "OFF" and "ON"
  • If persists: Do not use unit and call Blue Torch
Display reads:
   " E7"
CRC Self Check Error
  • Turn Unit "OFF" and "ON"
  • If persists: Do not use unit and call Blue Torch
Display reads:
   " E8"
Probe Tip Use Greater Than 8 Hours
  • Do not reuse probe tip
  • Replace with new probe tip
Display reads:
   " PL"
Stimulation Circuit Interrupted
  • Ensure all electrode are touching tissue
  • Ensure black lead is clipped to retractor
  • Check black lead connection to control unit
  • Clear alarm by pressing any button on probe tip
  • If persists, replace probe tip
Control Unit Not Responding to Probe Probe Handle Function
  • Try both buttons
  • Check cable connections
  • Turn Unit "OFF" then "ON", lights should flash on probe handle briefly
  • If persists, replace probe handle
Intermitent Stimulation Probe Handle Function
  • Do not hold down probe handle buttons
  • Check probe handle connections
No Tumescence Response When Expected System Not Functioning Properly
No Nerve
Penile Circumference is less than 8 cm
Patient's Current Threshold Low
  • Test unit on abdominal muscle to verify function
  • Using COARSE mode allow system to automatically increase current levels to 20 mA
  • Try another location for the Cavermosal nerve

Back to Top
 

 

Intended Use
The Blue Torch Medical Technologies CaverMap Surgical Aid is intended to provide stimulation to the body to locate and identify nerves and to test their excitability. The device is designed as an adjunct to the current open or laparoscopic prostatectomy and open colorectal procedures in which a nerve sparing technique is used. The Surgical Aid is not designed to replace the surgeon’s expertise in mapping out the neurovascular bundles.

CaverMap must be administered under a doctor's prescription to provide medical benefit.  CaverMap is often not eligible for a patient's insurance company reimbursement. An ABN Form is required to receive patient payments.